Why the study?
Does weighing QT intervals with the slope or amplitude of the T wave improve the reproducibility of the mean QT interval in post-myocardial infarction patients?
Does weighing QT intervals with the slope or amplitude of the T wave improve the reproducibility of the mean QT interval in post-myocardial infarction patients?
Weighing QT intervals based on the slope of the T wave significantly improves the reproducibility of the mean QT interval, primarily by reducing the negative impact of mismatch QT intervals.
Answers are grounded in this paper and cite their sources.
May support slope-weighted QT measurement in post-MI patients; leaves open validation against clinical endpoints.
OBJECTIVE: The reproducibility of QT interval measurements is low, even for the mean QT interval based on the standard ECG. In this study we analyzed whether the reproducibility of the mean weighed QT interval was better than the simple mean QT interval. The weighing was based on the amplitude of the T wave or the slope of the steepest tangent on the terminal part of the T wave. MATERIAL AND METHODS: 12-lead ECGs of 130 postmyocardial infarction patients were obtained. The QT intervals were measured by the tangent-method on two occasions by the same observer Mismatch QT intervals were defined as QT intervals that were measured at only one occasion. Sixteen ECGs were rejected. The data were split into 34 and 80 ECGs for optimization and validation of the weighing, respectively. The weighed QT dispersion was calculated as the weighed mean of the three longest minus the weighed mean of the three shortest QT intervals. RESULTS: Weighing with the slope increased the reproducibility by 41% (P = 3 10(-6)), but weighing with the amplitude reduced it by 20% (P = 0.02). However, if measurements with errors above 75 ms were rejected, weighing with the slope or the amplitude increased the reproducibility with 26% and 20% (P = 0.02), respectively. Weighing did not change the reproducibility of the weighed QT dispersion. CONCLUSION: Weighing with the slope improved the reproducibility of the mean weighed QT interval. However, if measurements with errors above 75 ms were rejected, weighing with the amplitude also increased the reproducibility. Weighing did not change the reproducibility of the weighed QT dispersion. Weighing is particularly efficient at reducing the negative impact of mismatch QT intervals on the reproducibility.
No takes yet. Share an insight, caveat, or question.
Lund et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: