Answers are grounded in this paper and cite their sources.
A single-shot coherent Raman imaging technique has been developed for spatially correlated one-dimensional high-fidelity gas-phase thermometry and multiplex chemical detection in flames. The technique utilizes two-beam phase matching, operating a single ultrashort pump/Stokes excitation pulse (7 fs) and a narrowband picosecond probe pulse (70 ps), interrogating a Raman active window of ∼4200 cm(-1) with ∼0.3 cm(-1) spectral resolution. The measurement geometry is formed intersecting the two beams shaped as laser-sheets and the one-coordinate spatial information is retrieved with a linespread function of <40 μm. The advance provides the possibility for the multiplexed measurement of all combustion relevant major species simultaneously with gaseous temperature monitored over a several millimeter field of view. The current technique is optimized in a premixed hydrocarbon flat-flame. At the flame-front, it is shown that direct imaging renders the temperature profile within ∼1% inaccuracy, whereas typical point-wise raster scanning may have relative systematic deviations up to 15% due to spatial averaging effects.
No takes yet. Share an insight, caveat, or question.
Bohlin et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: