Answers are grounded in this paper and cite their sources.
A commercial microwave acid-digestion system employing three types of closed vessels, a pressure-relief PFA-Teflon bomb (A), a pressure-relief Berghof all PTEE bomb (B) and a completely closed Parr microwave acid digestion bomb (C), was evaluated for sample digestion prior to the determination of trace elements in a biological tissue. The digestion procedure was based on sample dissolution with an HNO3-H2O2 mixture using microwave heating. The resulting solutions, analysed by flame and graphite furnace atomic absorption spectrometry, showed good agreement for all three closed-vessel techniques. Sample preparation time was approximately 10 min for vessels B and C, and 15 min for vessel A (including subsequent cooling time and preparation of the final solution).
No takes yet. Share an insight, caveat, or question.
Matusiewicz et al. (1989) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: