Although earmould venting is considered to make an earmould more comfortable, whether it does so remains to be proven in hearing-aid users. A hundred and six patients were sequentially fitted with a hearing aid in a randomized order with an unvented, a 0.8 mm vented or a 2 mm vented mould and questioned about various aspects of discomfort at the end of each trial period. The three moulds were equally uncomfortable (approximately 17% overall), moist (approximately 34% overall) and associated with an itchy sensation (approximately 36% overall). However, a 0.8 mm and a 2 mm vent significantly reduced the sensation of blockage from 35 to 22% (P less than 0.05) and 13% (P less than 0.001) respectively. A potential disadvantage of venting is feedback, but patients' reports of this can be confused with feedback due to incorrectly inserting the mould. A separate study showed that a 2 mm vent might cause feedback in approximately 10% of patients when used in conjunction with behind-the-ear aids whose maximum gain was 45 dB, but only when the gain was at 42 dB or greater.
No takes yet. Share an insight, caveat, or question.
MacKenzie et al. (1989) studied this question.