Why the study?
Does the Coherex FlatStent/FlatStent EF in-tunnel PFO closure system safely and effectively achieve PFO closure in patients with a clinical indication?
Does the Coherex FlatStent/FlatStent EF in-tunnel PFO closure system safely and effectively achieve PFO closure in patients with a clinical indication?
The second-generation Coherex FlatStent EF device demonstrated improved in-tunnel PFO closure rates (76% complete closure) compared to the first-generation device at 6 months, with an acceptable safety profile in a first-in-man study.
Answers are grounded in this paper and cite their sources.
May support second-generation device iteration for PFO closure; leaves open randomized confirmation of clinical benefit.
OBJECTIVE: The Coherex-EU Study evaluated the safety and efficacy of PFO closure utilizing novel in-tunnel PFO closure devices. BACKGROUND: Transcatheter closure of patent foramen ovale (PFO) followed the development of transcatheter closure devices designed to patch atrial septal defects (ASDs). The Coherex FlatStent™ and FlatStent™ EF devices were designed specifically to treat PFO anatomy. METHODS: A total of 95 patients with a clinical indication for PFO closure were enrolled in a prospective, multicenter first in man study at six clinical sites. Thirty-six patients received the first-generation FlatStent study device, and 57 patients received the second-generation FlatStent EF study device, which was modified based on clinical experience during the first 38 cases. Two patients enrolled to receive the first generation did not receive a device. RESULTS: At 6 months post-procedure, 45% (17/38) of the intention-to-treat (ITT) cohort receiving the first-generation FlatStent device had complete closure, 26% (10/38) had a trivial residual shunt, and 29% (11/38) had a moderate to large residual shunt. In the ITT cohort receiving the second-generation FlatStent EF device, 76% (43/57) had complete closure, 12% (7/57) had a trivial shunt, and 12% had a moderate to large shunt. Five major adverse events occurred, all without sequelae. CONCLUSION: This initial study of the Coherex FlatStent/FlatStent EF PFO Closure System demonstrated the potential for in-tunnel PFO closure. The in-tunnel Coherex FlatStent EF may offer an alternative to septal repair devices for PFO closure in appropriately selected patients; however, further investigation will be necessary to establish the best use of this device.
No takes yet. Share an insight, caveat, or question.
Sievert et al. (2012) studied this question.