OBJECTIVE: To compare a number of electronic tympanic, oral, axillary, and rectal measurements with those taken with a standard rectal mercury thermometer. DESIGN: Prospective open study. SETTING: County hospital, Denmark. SUBJECTS: 200 patients. INTERVENTIONS: Each of 200 patients had 6 electronic measurements of body temperature: 3 in the auditory canal using Ivac Core Check 2090A, Diateck 9000, and Genius 3000A, 1 in the axilla using Terumo Digital C202. 1 in the mouth using Terumo Digital C402, and 1 in the rectum using a Terumo Digital C402. These were compared with readings from a standard mercury glass thermometer in the rectum. MAIN OUTCOME MEASURES: Accuracy of electronic thermometry. RESULTS: The rectal electronic measurements were closest to the rectal mercury readings, with a mean (SD) of -0.05 degrees C (0.12), whereas the other measurements gave unacceptable SDs of temperature differences ranging from 0.41 degrees C to 0.53 degrees C. CONCLUSIONS: We conclude that electronic rectal temperature measurements are the most accurate. We do not recommend electronic tympanic, oral, or axillary measurements.
No takes yet. Share an insight, caveat, or question.
Bryce Jensen (2000) studied this question.