Why the study?
Does the Björk-Shiley tilting disc valve improve flow characteristics and reduce pressure gradients compared to other prostheses in an in vitro model?
Does the Björk-Shiley tilting disc valve improve flow characteristics and reduce pressure gradients compared to other prostheses in an in vitro model?
The Björk-Shiley tilting disc valve demonstrates excellent in vitro hemodynamic performance with the lowest pressure gradient among tested prostheses.
Answers are grounded in this paper and cite their sources.
In vitro hemodynamics favor the Björk-Shiley valve; leaves open translation to clinical outcomes or durability.
The mitral version of the Björk-Shiley tilting disc valve has an opening angle of only 50d` (B-S M) as compared to 60d` for the aortic valve (B-S A). Pulse duplicator studies with comparison of the performance of Starr-Edwards, Smeloff-Cutter, Kay-Shiley prostheses as well as a pig valve heterograft demonstrated the excellent flow characteristics of the Björk-Shiley tilting disc valves even in the modification for mitral valve replacement. As no gradient was found with a 27 mm mitral tilting disc valve (B-S M 27) at a comparatively high flow rate it is not necessary to introduce a larger valve from a haemodynamic point of view. A subaortic obstruction is also avoided with the tilting disc valve prosthesis, which showed the lowest gradient of all valves tested.
No takes yet. Share an insight, caveat, or question.
Björk et al. (1970) studied this question.