Why the study?
Does a stepwise approach with first-line mechanical polypropylene sheaths compared to laser-assisted sheaths improve outcomes and costs in patients undergoing transvenous lead extraction?
Does a stepwise approach with first-line mechanical polypropylene sheaths compared to laser-assisted sheaths improve outcomes and costs in patients undergoing transvenous lead extraction?
While laser-assisted transvenous lead extraction yields higher procedural success rates, a stepwise approach using mechanical polypropylene sheaths is more cost-effective and may be associated with fewer severe complications.
Answers are grounded in this paper and cite their sources.
May favor mechanical-first stepwise extraction for lower costs; leaves open need for randomized trials to confirm safety and efficacy.
AIMS: We sought to compare outcomes and costs of a stepwise approach to transvenous lead extraction (TLE) involving laser-assisted sheaths or mechanical polypropylene sheaths, with/without crossover. METHODS AND RESULTS: We prospectively included patients who underwent TLE (between August 2013 and December 2014) as part of a stepwise approach involving simple traction, lead snaring, and sheath-assisted dissection; all of these patients underwent a first-line polypropylene-sheath-extraction approach (Group A). The comparison group (Group B) was consecutive patients who had undergone TLE before August 2013, during which laser-assisted sheath extraction was the first-line approach. The number of patients in Group B was adjusted to match the number who eventually needed sheaths in Group A. Procedural data, outcomes, and costs were compared between groups (comparison of approaches) and in patients who needed sheath-assisted extraction (comparison of techniques). Overall, 521 leads were extracted (131 patients in Group A, 104 in Group B). Radiological and clinical success rates were similar; crossover from polypropylene to laser sheaths was needed in 10 patients in Group A (vs. none in Group B). Radiological (P< 0.001) and clinical (P= 0.01) success rates were higher and were achieved with a lower radiation exposure (P= 0.03) with laser sheaths in patients (60 in each group) who needed sheath-assisted extraction. Complication rates were similar in both groups (P= 0.66) but two deaths occurred in Group B. The laser approach had higher material cost (P= 0.002). CONCLUSIONS: Although laser-assisted TLE was more effective than polypropylene sheath-assisted TLE, the latter was associated with fewer complications and was more cost-effective.
No takes yet. Share an insight, caveat, or question.
Gaubert et al. (2016) studied this question.