In this paper, a highly sensitive methane (CH 4 ) and ammonia (NH 3 ) dual-gas photoacoustic spectroscopy (PAS) gas sensor based on a dual-path high-spot-density multipass cell (MPC) and dual-frequency photoacoustic cell (PAC) was demonstrated for the first time. A dual-path MPC mathematical model was established to achieve multiple independent rings and 223 excitations of dual gases. In order to integrate with the MPC more conveniently, the designed dual-frequency PAC was used to excite the Helmholtz resonance mode. A Raman fiber amplifier (RFA) and an erbium-doped fiber amplifier (EDFA) were employed to amplify the output power of the two diode lasers to enhance the excitation intensity. The experimental results indicated that when the output power was set as 50 and 300 mW, respectively, the 2 f signals for CH 4 and NH 3 detection with the high-spot-density MPC showed 41.91-fold and 41.30-fold enhancements compared to the system without MPC, demonstrating significant performance improvement. The minimum detection limits (MDLs) of CH 4 and NH 3 were found to be 65 and 374 ppb, respectively, when the average time of the CH 4 /NH 3 –PAS dual-gas sensor was 1000 s.