Why the study?
The optimal quantitative ablation index (AI) range for superior vena cava (SVC) electrical isolation during radiofrequency catheter ablation was unknown.
Does an optimized target ablation index (AI) value range of 350-400 improve the success rate of superior vena cava isolation in patients with atrial fibrillation?
Does an optimized target ablation index (AI) value range of 350-400 improve the success rate of superior vena cava isolation in patients with atrial fibrillation?
An ablation index target of 350-400 during high-power radiofrequency catheter ablation safely and effectively improves the success rate of superior vena cava isolation in patients with atrial fibrillation.
INTRODUCTION: To investigate the optimal range of quantitative ablation index (AI) value during superior vena cava (SVC) electrical isolation by radiofrequency catheter ablation (RFCA). METHODS: First, in a development cohort of patients with atrial fibrillation (AF), the RFCA with 40 W was performed to complete SVC isolation guided by the conduction breakthrough point from the right atrium to SVC. Then, the range of AI value was calculated by offline analysis on different segments of SVC. Lastly, for the validation of AF patients, the safety and effectiveness of SVC isolation with the optimized target range of AI value were evaluated with an additional adenosine test. RESULTS: A total of 101 patients with AF were included in the study (44 patients in the development cohort/57 in the validation cohort). The segmental ablation strategy was applied in 70% of the patients. According to the offline analysis of the AI values in the development cohort, the target AI value range was set as 350-400. The success rate of SVC isolation in the validation cohort was significantly higher than that in the exploration cohort (100% vs. 90.9%, p = .02), and no complications occurred in the exploration cohort. During the adenosine test, the recovery rate of electrical conduction in SVC was significantly lower than that in the pulmonary vein (3.5% vs. 17.5%). CONCLUSION: The target AI value with a range from 350 to 400 is safe and effective for high-power RFCA to complete SVC isolation.
No takes yet. Share an insight, caveat, or question.
Guan et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: