Answers are grounded in this paper and cite their sources.
The High energy Engineering X-ray (HEX) diffraction beamline at the National Synchrotron Light Source II (NSLS-II) at Brookhaven National Lab (BNL) is the first high-energy beamline capable of reaching 200 keV for a monochromatic beam. With the 3 GeV electron beam energy for the NSLS-II ring, only the superconducting wiggler (SCW) producing greater than 4 T peak field can cover these ranges with a sufficient number of photons. The 1.2 m-long HEX-SCW has a period length of 70 mm and a field strength on-axis of 4.3 T. It utilizes no liquid helium, and the vertical aperture size of the electron beam vacuum chamber is 8 mm. Unlike regular undulators/wigglers, there is no standard configuration for the magnetic measurement system for superconducting insertion devices. The NSLS-II Insertion Devices group has developed, in collaboration with the vacuum group, a novel in-vacuum Hall mapper with a 1.75 m in-vacuum linear motor and an in-vacuum flip coil system utilizing many commercial-off-the-shelf products. The measurements were conducted at the BNL, and the device was installed in the ring and commissioned. This paper provides a description of the SCW and its magnetic measurement systems, as well as a brief account of the installation and commissioning efforts.
No takes yet. Share an insight, caveat, or question.
Tanabe et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: