Precise quantification of human in vivo short echo time 1H spectra remains problematic at clinical field strengths due to broad peak linewidths and low signal-to-noise ratio (SNR). In this study, multiple STEAM spectra (TE = 20 ms, volume = 8 cm3) were acquired in a single individual at 1.5 T and 4 T to compare quantification precision. Test–retest STEAM spectra (volume = 1.5 cm3) were also acquired from the anterior cingulate and thalamus of 10 individuals at 4.0 T. Metabolite levels were quantified using automated software that incorporated field strength-specific prior knowledge. With the distinct methods of data acquisition, processing, and fitting used in this study, peak height SNR increased ∼80% while peak linewidth increased by ∼50% in the 8 cm3 volumes at 4.0 T compared to 1.5 T, resulting in an average increase in quantification precision of 39%. Metabolite levels from test–retest data (1.5 cm3 voxels at 4.0 T) were quantified with similar inter- and intraindividual variability. Magn Reson Med 44:185–192, 2000. Published 2000 Wiley-Liss, Inc.