We show the viability of high‐resolution two‐photon fluorescence imaging of fixed and live cells by exciting the fluorophores with a train of near‐infrared pulses with duration in the picosecond range. This is exemplified with a compact, diode‐pumped Nd:YVO 4 laser, emitting trains of 7‐ps pulses at a wavelength of 1064 nm, with a repetition rate of 200 MHz at two separate outputs. Incoherent combination of the outputs enabled two‐photon excitation with a repetition rate of 400 MHz. For a numerical aperture of 1.4 (oil), we used an average illumination power of up to 20–40 mW at the sample. The pulses were coupled into a beam scanning microscope, either directly or through a single mode glass fibre. Compared with standard femtosecond titanium–sapphire excitation conditions, our experiments were performed with a 2.5 or 5 times higher repetition rate, 30–70 times longer pulses and 10–35 times lower pulse peak intensity. The experiments indicate the possibility of significantly relaxing the temporal pulse width constraints for a series of applications.
No takes yet. Share an insight, caveat, or question.
Bewersdorf (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: