Key result
A novel pulsed field ablation system guided by real-time optical assessment showed that a mean tissue birefringence loss of 20.9% predicted durable lesions at 3 months, compared to 10.1% for reconnected areas (P<0.001).
Why the study?
Loss of bipolar electrograms immediately following pulsed field ablation makes lesion durability assessment challenging.
Does real-time tissue optical assessment using PS-OCR predict pulsed field lesion durability in patients with paroxysmal atrial fibrillation?
Does real-time tissue optical assessment using PS-OCR predict pulsed field lesion durability in patients with paroxysmal atrial fibrillation?
Real-time optical assessment of tissue birefringence drop during pulsed field ablation safely and accurately predicts lesion durability in patients with paroxysmal atrial fibrillation.
Answers are grounded in this paper and cite their sources.
Offers a potential real-time PFA durability marker where electrograms fail; leaves open whether guidance improves outcomes in larger trials.
AIMS: Loss of bipolar electrograms immediately after pulsed field ablation (PFA) makes lesion durability assessment challenging. OBJECTIVE: The aim of this trial (NCT06700226) was to evaluate a novel ablation system that can optically predict lesion durability by detecting structural changes in the tissue during ablation. METHODS AND RESULTS: Patients with paroxysmal atrial fibrillation underwent pulmonary vein isolation (PVI) using PFA (AblaView®, MedLumics). Using polarization-sensitive optical coherence reflectometry (PS-OCR), reflective characteristics of myocardial tissue and visualization of real-time contrast between healthy tissue and ablated tissue using a drop in tissue birefringence (BiR) was assessed. Wide antral PVI was performed using single point irrigated PFA (unipolar, 1800V, 3 trains, 21 s). Remapping was performed at 3 months. Primary efficacy outcome was the ability of PS-OCR to predict lesion durability at 3-month remapping. Serious adverse events were recorded. Ten patients were included. In total, 38/40 PVs could be isolated with the system. The mean drop of BiR was 17.3 ± 11.5%. Dragging across the ablation lines showed a persistent drop in BiR. During the remap procedures (8/10 patients ablated only with PFA), 12 PVs (37.5%) were found to be electrically reconnected. The mean loss of BiR during all PFA for durable lesions was 20.9%, while only 10.1% BiR loss was observed during the index ablation for reconnected areas (P < 0.001). None of the points with ≥17% loss of birefringence was found to be reconnected. CONCLUSION: This first-in-human study supports the use of real-time drop in tissue BiR for lesion assessment and durability during PFA delivery, and its procedural safety.
No takes yet. Share an insight, caveat, or question.
Martins et al. (2025) studied Paroxysmal atrial fibrillation (n=10). AblaView Unipolar PFA System was evaluated on Ability of PS-OCR to predict PFA lesion durability at 3-month remapping (mean loss of birefringence for durable vs reconnected lesions) (p=<0.001). A novel pulsed field ablation system guided by real-time optical assessment showed that a mean tissue birefringence loss of 20.9% predicted durable lesions at 3 months, compared to 10.1% for reconnected areas (P<0.001).