Answers are grounded in this paper and cite their sources.
We have fabricated and tested picovoltmeters in which the voltage source is connected in series with a calibrated resistor and a coil inductively coupled to a high-transition temperature superconducting quantum interference device operating at 77 K. The coil consists of either seven or ten turns of copper wire or of two turns patterned in a film of YBa2Cu3O7−x(YBCO). The lowest voltage noise achieved at 1 Hz was 30 pV Hz−1/2 and 2.3 pV Hz−1/2 for input coils made of copper and YBCO, respectively.
No takes yet. Share an insight, caveat, or question.
Miklich et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: