Why the study?
Limited data are available on the long-term efficacy of TAVR for BAV patients, whose valves are associated with elliptical openings and excess calcification.
Does simulated balloon-expandable TAVR result in worse mechanical metrics in patients with bicuspid aortic valves compared to trileaflet aortic valves?
Does simulated balloon-expandable TAVR result in worse mechanical metrics in patients with bicuspid aortic valves compared to trileaflet aortic valves?
Simulated balloon-expandable TAVR in bicuspid aortic valves demonstrates worse mechanical metrics, including higher leaflet stress and lower expansion ratios, compared to trileaflet valves, which correlates with calcium volume.
Hypothesis-generating for BAV TAVR mechanics; leaves open clinical translation pending human outcome data.
BACKGROUND: Bicuspid aortic valve (BAV) disease is the most common congenital heart defect and is associated with elliptical openings and excess calcification. Limited data are available on long-term efficacy of transcatheter aortic valve replacement (TAVR) for BAV patients. OBJECTIVES: The purpose of this study was to compare simulated mechanical metrics post-TAVR between BAV and trileaflet patients with calcific aortic stenosis. METHODS: Pre-TAVR computed tomography scans of trileaflet (n = 22) and BAV (n = 25) were used to segment the aortic valve. Simulated balloon-expandable stent deployment was performed from a previously developed protocol. The bioprosthetic leaflets were then pressurized under a physiological pressure. Stent waist expansion ratio, aspect ratio (AR), leaflet systolic normalized geometric orifice area (GOA), and maximum leaflet diastolic stress were measured. RESULTS: BAVs had lower waist expansion ratio (87.8% ± 2.6% vs 89.4% ± 1.4%; P < 0.001), lower waist AR (94.6% ± 2.8% vs 98.2% ± 1.0%; P < 0.001), lower normalized GOA (88.9% ± 0.8% vs 89.5% ± 0.7%; P < 0.01), and higher maximum stress (2.81 ± 0.44 vs 2.29 ± 0.25 MPa; P < 0.001) on average compared to trileaflet patients, respectively. Within the BAV group, calcium volume was correlated with waist AR (r = -0.75, P < 0.001), normalized GOA (r = -0.68, P < 0.001), and maximum stress (r = 0.66, P < 0.001). CONCLUSIONS: Balloon-expandable TAVR for BAV has worse simulated mechanical metrics compared to TAVR for trileaflet aortic valve patients and is associated with high calcium volume.
No takes yet. Share an insight, caveat, or question.
Yeats et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: