Why the study?
Optimum filter selection for pre-processing camera-based remote photoplethysmography data in signal conditioning is still not straightforward.
Chebyshev II filters (2nd, 4th, and 6th order) provide the optimal signal-to-noise ratio for pre-processing remote photoplethysmography data.
Guides filter selection in rPPG pipelines; leaves open clinical validation in cardiovascular cohorts.
Remote photoplethysmography (rPPG) using camera-based imaging has shown excellent potential recently in vital signs monitoring due to its contactless nature. However, the optimum filter selection for pre-processing rPPG data in signal conditioning is still not straightforward. The best algorithm selection improves the signal-to-noise ratio (SNR) and therefore improves the accuracy of the recognition and classification of vital signs. We recorded more than 300 temporal rPPG signals where the noise was not motion-induced. Then, we investigated the best digital filter in pre-processing temporal rPPG data and compared the performances of 10 filters with 10 orders each (i.e., a total of 100 filters). The performances are assessed using a signal quality metric on three levels. The quality of the raw signals was classified under three categories; Q1 being the best and Q3 being the worst. The results are presented in SNR scores, which show that the Chebyshev II orders of 2nd, 4th, and 6th perform the best for denoising rPPG signals.
No takes yet. Share an insight, caveat, or question.
Guler et al. (2023) studied this question.