This is the authors' abstract. We don't add key points for this paper.
This article investigates decoupled motion control and dimension optimization of composite notched continuum mechanisms. In general, the end-effectors of endoscopic surgical robots predominantly consist of rigid articulated actuators, which exhibit limited maneuverability and face challenges in constrained operational environments. The introduction of continuum mechanisms has emerged as a key solution to address these limitations. In this article, the design, analysis, and development of a novel six-degrees-of-freedom (6-DOF) composite continuum surgical robot are presented. Kinematic modeling of the continuum mechanisms is performed, and a decoupled kinematic model of the composite continuum mechanisms is constructed. Furthermore, based on the local and global dexterity of the composite continuum mechanisms, the optimization of the two-segment lengths of the composite continuum mechanisms is completed. Subsequently, both the forward and inverse kinematic models of the 6-DOF composite continuum surgical robot are developed. Finally, through a series of motion control experiments, the decoupled kinematic model of the prototype is proved to be effective. The prototype has a certain load capacity and can accomplish simple trajectory planning motion, which has the potential application in the field of single-hole interventional minimally invasive surgery.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: