This study presents a high-precision method, using a preconcentrator–gas chromatograph with microelectron capture detector (GC- μ ECD), to measure SF 6 at ambient levels. Carboxen 1000 was used as an adsorbent for the preconcentrator and exhibited a high adsorption efficiency for N 2 O and SF 6 and low adsorption efficiency for O 2 . This enabled the selective removal of atmospheric O 2 from analytes and improved repeatability of the SF 6 peak that followed the O 2 peak, in a separation column of activated alumina F1. In addition, the increased sensitivity resulting from preconcentrated SF 6 improved the signal-to-noise ratio. This led to better analytical precision in comparison with other measurement methods including the conventional and forecut–backflush (FCBF) methods. The precision-to-drift ratios of the conventional, FCBF, and preconcentration methods were 0.11, 0.10, and 0.03, respectively. Analytical precision of the preconcentration method was 0.08% for 10 consecutive injections; this was the best among the three methods. The long-term drift of the SF 6 response was inversely proportional to the laboratory pressure. Based on this finding, room pressure can be used to correct for ECD signal drift, with an uncertainty of 0.14% over a 48-h period, using the preconcentration method. Another advantage of the preconcentration method was the excellent linearity of the SF 6 response to a wide range of concentrations, including its ambient concentration.