Key result
A novel 3D printed simulator for mitral valve transcatheter edge-to-edge repair was successfully tested and found feasible for practicing procedural steps and compatible with clinical fluoroscopy.
Why the study?
MitraClip training is technically challenging with a steep learning curve, and existing simplified models lack anatomical fidelity, realistic materials, and procedural scenarios.
A novel 3D printed simulator for TEER was successfully developed and demonstrated feasibility for procedural training and compatibility with fluoroscopy.
Animal feasibility supports 3D-printed TEER simulator refinement; clinical training benefit remains unproven and requires prospective validation.
Mitral regurgitation is a common valvular disorder. Transcatheter edge-to-edge repair (TEER) is a minimally invasive technique which involves holding together the middle segments of the mitral valve leaflets, thereby reducing regurgitation. To date, MitraClip™ is the only Food and Drug Administration (FDA)-approved device for TEER. The MitraClip procedure is technically challenging, characterised by a steep learning curve. Training is generally performed on simplified models, which do not emphasise anatomical features, realistic materials, or procedural scenarios. The aim of this study is to propose a novel, 3D printed simulator, with a major focus on reproducing the anatomy and plasticity of all areas of the heart involved and specifically the ones of the mitral valve apparatus. A three-dimensional digital model of a heart was generated by segmenting computed tomography (CT). The model was subsequently modified for: (i) adding anatomical features not fully visible with CT; (ii) adapting the model to interact with the MitraClip procedural equipment; and (iii) ensuring modularity of the system. The model was manufactured with a Polyjet technology printer, with a differentiated material assignment among its portions. Polypropylene threads were stitched to replicate chordae tendineae. The proposed system was successfully tested with MitraClip equipment. The simulator was assessed to be feasible to practice in a realistic fashion, different procedural aspects including access, navigation, catheter steering, and leaflets grasping. In addition, the model was found to be compatible with clinical procedural imaging fluoroscopy equipment. Future studies will assess the effect of the proposed training system on improving TEER training.
No takes yet. Share an insight, caveat, or question.
Bertolini et al. (2022) studied Mitral regurgitation. Novel 3D printed simulator for TEER was evaluated on Feasibility of practicing procedural aspects and compatibility with clinical procedural imaging fluoroscopy equipment. A novel 3D printed simulator for mitral valve transcatheter edge-to-edge repair was successfully tested and found feasible for practicing procedural steps and compatible with clinical fluoroscopy.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: