Answers are grounded in this paper and cite their sources.
Instrumentation study demonstrates separation of low-energy ions in Earth's ionosphere, suggesting enhanced dynamic range and mass resolution without foil scattering effects.
When incorporated into a top-hat electrostatic analyzer, a gate electrode enables the separation of ions by their mass-per-charge with modest mass resolution (M/∆M ∼ 10). Gated-time-of-flight (TOF) instruments avoid the energy straggling and angular scattering effects prevalent in foil-based detection systems, providing more pristine measurements of three-dimensional distribution functions of incident ions. Gated-TOF implementations are ideal for measuring the properties of low-energy (i.e., <100 eV) thermal ions in various space environments. We present an instrument prototype capable of separating H+, He+, O+, and O2+ in Earth's ionosphere and demonstrate that in addition to providing species determination, precise operation of the gate electrode provides an electronically adjustable geometric factor that can extend a single instrument's dynamic range by several orders of magnitude.
No takes yet. Share an insight, caveat, or question.
Gershman et al. (2023) studied this question.