High-frequency irreversible electroporation (H-FIRE), which employs high-frequency bipolar pulses (HFBPs) with high voltage has been successfully used for tumor ablation with the inhibition of muscle contraction. This article designs an undershoot damping circuit with magnetic isolation drive embedded (MI-UD) in main topology of step-up transformers to reduce the undershoot voltage resulting from the energy release of excitation inductance and improve the pulse frequency when outputting HFBPs. The proposed main circuit topology could generate HFBPs with one magnetic core, and no additional magnetic core reset circuit is needed for the excitation inductance. Moreover, the designed drive circuit only needs two control signals with delayed turn-off, including the controls of the MI-UD circuit. Theoretical analysis, simulation, and experimental test indicate that three modules of step-up pulse transformer could output HFBPs. The pulse amplitude could reach ±3 kV with the voltage gain of 11.1 (ratio of 1:4, three modules) and dc supply of 270 V. The pulsewidth ranges from 300 ns to 2μs. A total energized time of 100μs within the burst can be achieved to meet the H-FIRE treatment. The maximum frequency is 400 kHz within the burst. The undershoot at 2μs and ±3 kV is only 3.3%.