An instrument for simultaneous measurement of foveal spectral reflectance and cone-photoreceptor directionality is described. The key element is an imaging spectrograph (spectral range of 420-790 nm) with its entrance slit conjugate to the pupil plane of a human eye. A 1.9-deg spot on the retina is sampled in 1 s. Video observation of the retina and the pupil facilitates proper alignment. Measurements were performed on 21 healthy subjects. Model analysis of the spectra provided densities of photostable ocular absorbers. As an example, macular pigment and melanin are discussed in more detail. Spatial profiles exhibited the optical Stiles-Crawford effect, reflecting cone-photoreceptor directionality.
No takes yet. Share an insight, caveat, or question.
Zagers et al. (2002) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: