Objective: This study aimed to evaluate the accuracy of the portable Somt (Compumedics Limited, Melbourne, Australia) sleep data acquisition device against the current gold-standard of technician-attended, laboratory-based polysomnography (PSG) for the diagnosis of sleep disordered breathing (SDB). Methods: Patients attending the sleep disorders laboratory for overnight PSG for the evaluation of suspected SDB were fitted with both the Somt device and a PSG system (Compumedics E-Series). An apnea-hypopnea index (AHI) was calculated after manual scoring of respiratory events recorded using the Somt and PSG systems. Results: 37 patients (29 male, 8 female) participated in the study. The mean difference between the AHI estimated by the two devices showed good agreement, with the Somt device AHI being similar to that obtained by PSG. Mean difference (PSG -Somt AHI), -0.5 events per hour (95% CI -4.4 to 5.4, p=0.83). The Somt device was able to correctly classify the presence or absence of SDB (defined as PSG AHI e"15) with a sensitivity of 96%, specificity of 83%, positive predictive value of 92% and a negative predictive value of 91%.
No takes yet. Share an insight, caveat, or question.
Cunnington et al. (2009) studied this question.