Answers are grounded in this paper and cite their sources.
In this study, the optimized structure of a femoral prosthesis for a patient was determined by biomechanical analysis, and the materials that match the model of the prosthesis were determined by multi-objective comprehensive evaluation using a fuzzy matter-element method. The suitable material for the ellipsoidal femoral head of the prosthesis was determined to be a carbon-fiber-enhanced polyether-ketone (CF/PEK) composite, and that for the stem was determined to be a zirconium–niobium alloy (Zr–Nb alloy). The study successfully demonstrated the potential of the developed method for use in selecting the best structure and materials for fabricating a customized prosthesis.
No takes yet. Share an insight, caveat, or question.
Monan Wang (2016) studied this question.