In the vast panorama of optical-based detection strategies, photoacoustic spectroscopy is particularly suitable for its versatility, high sensitivity, and scalability potential. In cantilever-enhanced photoacoustic sensors, the interferometric readout of the membrane oscillation guarantees exceptional signal-to-noise ratio, often at the expense of limited compactness. In this work, we propose a novel approach combining cantilever-enhanced photoacoustic trace-gas excitation with self-mixing interferometry for absorption signal readout. The sensor performance has been compared in terms of detection sensitivity and stability over time to that achieved with a more complex state-of-the-art readout system using a bulky balanced Michelson interferometer. Both sensors, operated with the same excitation laser emitting at 4.57 µm and addressing the same target N 2 O line, demonstrated the same spectroscopic results in terms of signal-to-noise ratio of the acquired spectra and the same minimum detection limit of 90 parts-per-billion at tens of seconds integration time. The self-mixing readout benefits from a much reduced size, paving the way for future system downsizing and easier integration while maintaining high-sensitivity detection levels. Moreover, the intrinsic wavelength independence of photoacoustic spectroscopy, together with the broad-spectral adaptability of self-mixing readout, allows, in principle, a wide wavelength-tailorability of the sensor.
No takes yet. Share an insight, caveat, or question.
Gabbrielli et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: