Answers are grounded in this paper and cite their sources.
The aim of this study was to reproduce relaxivity‐iron calibration in hepatic iron overload using a Monte Carlo model, and further extend the model with multiple spin echo (MSE) imaging. As previously reported, relationships between relaxation rates ( and single spin echo R 2 ) and liver iron concentration (LIC) can be characterized by a Monte Carlo model incorporating realistic liver structure, iron distribution, and proton mobility. In this study, relaxivity‐iron calibration curves at 1.5 and 3.0 T were simulated using the Monte Carlo model. Furthermore, the model was extended with MSE imaging, and iron calibrations were evaluated using two different fitting models: mononexponential with a constant offset and nonmonoexponential. Results consistent with previous empirical calibrations and Monte Carlo predictions were accurately reproduced for relaxivity‐iron calibration. The predicted and single spin echo R 2 increased by a factor of 2.00 and 1.51, respectively, at 1.5 versus 3.0 T. MSE signals and their corresponding R 2 depended strongly on LIC, interecho time, and field strength. Preliminary results showed that a nonmonoexponential model accurately characterizes the simulated MSE signals, and that strong correlations were found between predicted relaxation parameters and LIC. In conclusion, relaxivity‐iron calibration is reproducible using the proposed Monte Carlo model. Furthermore, this model can be readily extended to other important applications, including predicting signal behavior for MSE imaging.