Answers are grounded in this paper and cite their sources.
The accident at the Tokyo Electric Power Company’s (TEPCO’s) Fukushima Daiichi Nuclear Power Plant in March 2011 elevated the concerns of many citizens regarding radiation exposure. Therefore, a dosimeter that enables anyone to easily measure radiation exposure was developed for use by residents in the event of a nuclear accident (D-Shuttle; Chiyoda Technol Co., Tokyo, Japan). The D-Shuttle is widely used, mainly in the Fukushima Prefecture; however, there are few user evaluations on its basic performance. This paper investigates the basic performance of the D-Shuttle, including the stability of the measured values. It was found that there was no significant sensitivity difference among the multiple D-Shuttle units or in the measured values before and after battery replacement. This indicates the stability of D-Shuttle measurements. As α- and β-rays could not be detected, D-Shuttle should only be used for γ-ray measurements. The measured value on the back side of the D-Shuttle was 66.7% of that on the front side. This indicates that the D-Shuttle is less sensitive on the back side.
No takes yet. Share an insight, caveat, or question.
Akisawa et al. (2025) studied this question.