This is the authors' abstract. We don't add key points for this paper.
In this paper, we report high-precision absolute distance and vibration measurements performed with frequency scanned interferometry. Absolute distance was determined by counting the interference fringes produced while scanning the laser frequency. High-finesse Fabry-Perot interferometers were used to determine frequency changes during scanning. A dual-laser scanning technique was used to cancel drift errors to improve the absolute distance measurement precision. A new dual-channel FSI demonstration system is also presented which is an interim stage toward practical application of multi-channel distance measurement. Under realistic conditions, a precision of 0.3 microns was achieved for an absolute distance of 0.57 meters. A possible optical alignment system for a silicon tracker is also presented.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2011) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: