Answers are grounded in this paper and cite their sources.
We have characterized a full-scale transition-edge sensor (TES) microcalorimeter array as a milestone towards the demonstration of technology readiness level (TRL) 5 for the detector array for the X-ray Integral Field Unit (X-IFU) instrument on ESA's future flagship X-Ray Observatory called Athena. We fabricated a 90 mm full-scale prototype TES array and measured the properties and the performance in a newly developed platform in which up to 960 out of ∼3,200 pixels can be read out. In this paper, we report on measurements of the uniformity of the transition temperature and shape, the uniformity of the spectral energy resolutions at 7 keV and 10 keV, the thermal crosstalk for the first, the diagonal, and the second nearest neighbors, and the energy sensitivities to environment perturbations for the magnetic field, the TES bias voltage, and the heat bath temperature.
No takes yet. Share an insight, caveat, or question.
Sakai et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: