Key result
A low-volume plasma renin activity ELISA requiring only 100 μL of plasma was successfully validated according to FDA guidelines, demonstrating non-inferiority to established radioimmunoassays.
Why the study?
Investigations of plasma renin activity in children require small-volume, regulatory compliant bioanalytical assays that address paediatric ethical constraints.
A low-volume PRA ELISA was successfully validated for pediatric use, enabling regulatory-compliant determination of plasma renin activity in neonates with heart failure using only 100 μL of plasma.
May facilitate PRA testing in neonates with limited blood volume; leaves open prospective outcome validation in pediatric heart failure.
OBJECTIVES: Investigations of plasma renin activity (PRA) in children are urgently required. Small-volume, regulatory guideline compliant, bioanalytical assays tailored for paediatric application could facilitate to overcome this hurdle. Ethical constraints given e.g. by the European Medicines Agency need to be addressed and reliable data generation in line with Good Clinical Laboratory Practice must be ensured. METHODS: A PRA enzyme-linked immunosorbent assay (ELISA) was tailored for paediatric application and validated in the context of the U.S. Food and Drug Administration bioanalytical guideline. Performance verification of the assay was conducted by participation in an interlaboratory ring test, evaluation of incurred sample reanalysis and an application-orientated approach in children. RESULTS: A five-fold reduction of required plasma volume to 100 μL was achieved without limiting the calibration range. Between-run accuracy and precision varied no more than 5.0% and 6.3%, respectively. No substantial matrix effect was detected and the inter-run precision for parallelism was 11.1%. Stability experiments approved the freeze-thaw stability, short-term stability as well as 37 weeks of long-term stability. The assay successfully participated in the interlaboratory ring test, showing non-inferiority regarding radioimmunoassay (RIA). Moreover, PRA in plasma samples of neonates was successfully determined. Conducted incurred sample reanalysis confirmed the comparability and reliability of the assay with regard to international regulatory bioanalytical guidelines. CONCLUSION: A fit-for-purpose PRA ELISA characterised by low-volume application was successfully established, indicating non-inferiority regarding commonly applied RIAs. Reliability of the regulatory-compliant PRA assay was proven by participation in an interlaboratory ring test and its application in a paediatric population.
No takes yet. Share an insight, caveat, or question.
Suessenbach et al. (2019) studied Heart failure. Low-volume plasma renin activity ELISA vs. Radioimmunoassay (RIA) was evaluated on Assay validity (accuracy, precision, stability) according to FDA bioanalytical guidelines. A low-volume plasma renin activity ELISA requiring only 100 μL of plasma was successfully validated according to FDA guidelines, demonstrating non-inferiority to established radioimmunoassays.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: