Answers are grounded in this paper and cite their sources.
OBJECTIVE: To determine if using an N95 filtering face-piece respirator concurrently with a loose-fitting powered air-purifying respirator (PAPR) offers additional protection to the wearer. METHODS: We used a breathing mannequin programmed to deliver minute volumes of 25 L/min and 40 L/min. We measured the baseline protection factor of the PAPR with its motor operational and then deactivated (to simulate mechanical or battery failure). We tested 3 replicates of 3 different N95 models. We glued each N95 to the breathing mannequin and obtained a minimum protection factor of 100 at 25 L/min. We then placed the PAPR on the mannequin and took protection factor measurements with the N95-plus-PAPR combination, at 25 L/min and 40 L/min, with the PAPR operational and then deactivated. RESULTS: The N95 significantly increased the PAPR's protection factor, even with the PAPR deactivated. The effect was multiplicative, not merely additive. CONCLUSIONS: An N95 decreases the concentration of airborne particles inspired by the wearer of a PAPR.
No takes yet. Share an insight, caveat, or question.
Roberge et al. (2008) studied this question.