Key result
All 8 participating centers correctly interpreted whole-mount electron microscopy findings for normal and dense-granule-deficient platelets in a pilot external quality assessment survey.
Platelet electron microscopy for dense-granule deficiency can be successfully assessed by external quality assessment programs.
This is the authors' abstract. We don't add key points for this paper.
Performance on specialized diagnostic tests for platelet disorders, including dense-granule deficiency, is rarely evaluated by external quality assessment (EQA). Members of the North American Specialized Coagulation Laboratory Association that evaluate platelet dense-granule deficiency commonly use whole-mount electron microscopy (EM) methods. This observation led us to develop a pilot EQA survey with standardized EM images and clinical samples on grids from a healthy control subject and a subject with dense-granule deficiency. The survey participants were 8 centers, including 2 with no experience in platelet whole mount EM. All participants, including inexperienced sites, correctly interpreted findings for the normal and dense-granule-deficient platelets. Among experienced sites, agreement was excellent (>82%) on platelet structures to count or not count as dense granules. Participants indicated that future EQA challenges should include clinical samples on grids and standardized images. This is the first report that platelet EM can be assessed by EQA.
Answers are grounded in this paper and cite their sources.
Supports pilot external quality assessment for platelet electron microscopy; leaves open wider validation before routine adoption.
No takes yet. Share an insight, caveat, or question.
Hayward et al. (2009) studied Platelet dense-granule deficiency (n=8). External quality assessment (EQA) survey was evaluated on Correct interpretation of findings for normal and dense-granule-deficient platelets. All 8 participating centers correctly interpreted whole-mount electron microscopy findings for normal and dense-granule-deficient platelets in a pilot external quality assessment survey.
Synapse has enriched 10 related papers. Consider them for comparative context: