Key result
Radiofrequency energy delivered through a 3830-pacemaker lead in chicken breasts produced maximum lesion areas of 0.78 cm2 at 10W (direct method) and 0.72 cm2 at 50W (close bipolar method).
Why the study?
Because radiofrequency catheters must deliver energy close to arrhythmia foci, an approach to deliver energy to deeper locations using a pacemaker lead was explored.
Does delivering RF energy through a pacemaker lead create adequate lesions in a preclinical tissue model?
Does delivering RF energy through a pacemaker lead create adequate lesions in a preclinical tissue model?
Delivering RF energy through a standard pacemaker lead is feasible in a tissue model, suggesting a potential novel approach for deep septal or epicardial ablation.
Ex vivo feasibility cautions against clinical adoption; leaves open investigation of lead-based ablation for deep targets.
Introduction Radiofrequency (RF) is the preferred thermal energy used in electrophysiology. RF catheter must deliver the energy close to arrhythmia foci. A new method to deliver RF to deeper locations using a pacemaker lead is explored. Methods A Medtronic 3830 lead screwed in chicken breasts delivered 50 watts RF energy in three methods: A) direct fashion (RF catheter touching the proximal end of the 3830 lead, acting as an extension of RF catheter), or B) 3830 lead as a return patch (RF delivered in the bath without contact), or C) 3830 lead as a return patch (RF delivered touching the breast surface close to the 3830 lead screwed deep in the flesh). Different power settings were also tested. Lesion surface area is reported in cm2. Results 76 measurements were available. Bigger lesions were obtained at 10W method A (0.78cm2), 50W method C (0.72cm2) and 5W method B (0.44cm2). High impedances were noted at 10W and 50W with tissue remaining attached to the lead when removed. Conclusion RF can be delivered to deeper foci through a 3830-pacemaker lead with maximum size lesion formation using proximal unipolar direct delivery and proximal close bipolar as the return patch. In humans, it opens a path to attain deep septal foci (LV summit) or epicardial structures (vein of Marshall, transmural ablation from RF endocardial to LV coronary sinus lead as return patch): using standard, 4F pacemaker leads, and 2F small EP catheters or even isolated guidewires.
No takes yet. Share an insight, caveat, or question.
AYALA-VALANI et al. (2022) studied this question. Radiofrequency ablation via Medtronic 3830 pacemaker lead vs. Different delivery methods and power settings was evaluated on Lesion surface area (cm2). Radiofrequency energy delivered through a 3830-pacemaker lead in chicken breasts produced maximum lesion areas of 0.78 cm2 at 10W (direct method) and 0.72 cm2 at 50W (close bipolar method).