A high-throughput broadband (1–24-Å) x-ray spectrometer has been demonstrated on the divertor injection tokamak experiment (DITE) tokamak. A hexagonal rotor supporting six diffractors may be driven in several modes, ranging from a full spectral survey at ∼10 Hz to a stationary, monochromator mode. Wavelength resolution, 500≲λ/Δλ≲1000, is governed by gridded or slotted collimators. A multiwire gas proportional counter provides a measure of energy discrimination, which together with the large instrument aperture, gives sufficient sensitivity and signal/noise ratio to allow measurement of the continuum radiation from the tokamak. The instrument has a self-contained vacuum system which allows full spatial scans of the DITE plasma. Data acquisition and drive mechanisms for the rotor and filter selection, are controlled remotely from a computer. Results are presented of fast spectral surveys and time evolution of impurity emission during impurity injection.
No takes yet. Share an insight, caveat, or question.
Barnsley et al. (1986) studied this question.