Key result
The third-generation Vigileo-FloTrac device accurately tracked cardiac output changes during preload changes but not during phenylephrine or ephedrine administration.
Why the study?
Does the third-generation Vigileo-FloTrac system accurately track changes in cardiac output induced by vasopressors and preload changes compared to esophageal Doppler in anesthetized patients?
Does the third-generation Vigileo-FloTrac system accurately track changes in cardiac output induced by vasopressors and preload changes compared to esophageal Doppler in anesthetized patients?
The third-generation Vigileo-FloTrac system accurately tracks cardiac output changes from preload variations but is inaccurate during vasopressor (phenylephrine, ephedrine) administration.
Answers are grounded in this paper and cite their sources.
Caution warranted during vasopressor use with Vigileo-FloTrac; observational data leaves open need for randomized validation.
BACKGROUND: Cardiac output (CO) monitoring based on pulse contour analysis (Vigileo-FloTrac) has the potential to be used for goal-directed fluid therapy in the perioperative setting. However, factors such as vasopressor usage may impact Vigileo-FloTrac's reliability in tracking CO changes. We tested third-generation Vigileo-FloTrac system's ability to accurately measure the changes in CO induced by vasopressor administration and increased preload in comparison with esophageal Doppler measurements. METHODS: In 33 anesthetized patients, CO was monitored simultaneously by the third-generation Vigileo-FloTrac and esophageal Doppler. Hemodynamic challenges included phenylephrine (to increase vasomotor tone), ephedrine (to increase myocardial contractility and heart rate), and whole-body tilting (to increase preload). Measurements were performed before and after each intervention. RESULTS: Overall, 176 pairs of CO measurements were obtained. The difference between paired pulse contour and Doppler measurements of CO was 0.14 ± 2.13 L/min (mean ± SD), and the percentage error (2 SD of the difference divided by the mean CO of the reference method) was 66%. The trending ability of pulse contour versus Doppler was 23% (concordance, the percentage of the total number of data points that are in 1 of the 2 quadrants of agreement) after phenylephrine treatment, 69% (concordance) after ephedrine treatment, and 96% (concordance) after whole-body tilting. CONCLUSIONS: The pulse contour method of measuring CO, as implemented in the third-generation Vigileo-FloTrac device, accurately tracks changes in CO when preload changes. However, the pulse contour method does not accurately track changes in CO induced with phenylephrine and ephedrine.
No takes yet. Share an insight, caveat, or question.
Meng et al. (2011) studied Anesthetized patients (n=33). Third-generation Vigileo-FloTrac system vs. Esophageal Doppler was evaluated on Difference between paired pulse contour and Doppler measurements of cardiac output (MD 0.14 L/min). The third-generation Vigileo-FloTrac device accurately tracked cardiac output changes during preload changes but not during phenylephrine or ephedrine administration.