Key result
The phaser adiabatic T2-preparation module significantly reduced chest fat SNR by 53% and increased right coronary artery vessel sharpness by 24% compared to conventional adiabatic T2-preparation with fat saturation.
Why the study?
Existing methods lacked a T2 preparation module capable of simultaneously providing robust fat suppression and efficient T2 preparation without requiring an additional fat suppression module for T2-weighted imaging at 3T.
Does the phaser adiabatic T2-Prep module improve fat suppression and vessel sharpness compared to conventional T2-Prep modules in cardiac MRI at 3T?
Does the phaser adiabatic T2-Prep module improve fat suppression and vessel sharpness compared to conventional T2-Prep modules in cardiac MRI at 3T?
Answers are grounded in this paper and cite their sources.
The novel phaser adiabatic T2-preparation module provides robust fat suppression and increases coronary vessel sharpness at 3T MRI without compromising T2 preparation capabilities.
Purpose: Designing a new T2 preparation (T2-Prep) module in order to simultaneously provide robust fat suppression and efficient T2 preparation without requiring an additional fat suppression module for T2-weighted imaging at 3T. Methods: The tip-down RF pulse of an adiabatic T2 preparation (T2-Prep) module was replaced by a custom-designed RF excitation pulse that induces a phase difference between water and fat, resulting in a simultaneous T2 preparation of water signals and the suppression of fat signals at the end of the module (now called a phaser adiabatic T2-Prep). Using numerical simulations, in vitro and in vivo ECG-triggered navigator gated acquisitions of the human heart, the blood, myocardium and fat signal-to-noise ratio and right coronary artery (RCA) vessel sharpness using this approach were compared against previously published conventional adiabatic T2-Prep approaches Results: Numerical simulations predicted an increased fat suppression bandwidth and decreased sensitivity against transmit magnetic field inhomogeneities using the proposed approach, while preserving the water T2 preparation capabilities. This was confirmed by the tissue signals acquired on the phantom and the in vivo MRA, which show similar blood and myocardium SNR and CNR and significantly reduced fat SNR compared to the other methods tested. As a result, the RCA conspicuity was significantly increased and the motion artifacts were visually decreased. Conclusion: A novel fat-suppressing T2-preparation method was developed and implemented that demonstrated robust fat suppression and increased vessel sharpness compared with conventional techniques, while preserving its T2 preparation capabilities.
No takes yet. Share an insight, caveat, or question.
Arn et al. (2020) studied Healthy volunteers (n=6). Phaser adiabatic T2-preparation (PA-T2-Prep) vs. Conventional adiabatic T2-Prep (CA-T2-Prep) with and without fat saturation was evaluated on Chest fat SNR reduction (53% reduction, p=0.03). The phaser adiabatic T2-preparation module significantly reduced chest fat SNR by 53% and increased right coronary artery vessel sharpness by 24% compared to conventional adiabatic T2-preparation with fat saturation.