Key result
Image-based navigation for dual-phase 3D whole-heart MRI significantly reduced mean scan time by 3 minutes and 16 seconds compared to diaphragmatic navigation in patients with congenital heart disease, while maintaining similar image quality.
Why the study?
Does image-navigation (iNAV) reduce acquisition time and improve image quality compared to diaphragmatic-navigation (dNAV) in dual-phase whole-heart MRI for congenital heart disease?
Cross-Sectional (n=49)
Blinded image analysis
Randomized order of scan acquisition
No
Does image-navigation (iNAV) reduce acquisition time and improve image quality compared to diaphragmatic-navigation (dNAV) in dual-phase whole-heart MRI for congenital heart disease?
Image-navigation for dual-phase whole-heart MRI significantly reduces scan time while maintaining or improving diagnostic image quality compared to conventional diaphragmatic navigation in patients with congenital heart disease.
Answers are grounded in this paper and cite their sources.
May support shorter dual-phase whole-heart MRI scans in congenital heart disease without quality loss; leaves open randomized confirmation before practice change.
BACKGROUND: Dual-phase 3-dimensional whole-heart acquisition allows simultaneous imaging during systole and diastole. Respiratory navigator gating and tracking of the diaphragm is used with limited accuracy. Prolonged scan time is common, and navigation often fails in patients with erratic breathing. Image-navigation (iNAV) tracks movement of the heart itself and is feasible in single phase whole heart imaging. To evaluate its diagnostic ability in congenital heart disease, we sought to apply iNAV to dual-phase sequencing. METHODS: Healthy volunteers and patients with congenital heart disease underwent dual-phase imaging using the conventional diaphragmatic-navigation (dNAV) and iNAV. Acquisition time was recorded and image quality assessed. Sharpness and length of the right coronary (RCA), left anterior descending (LAD), and circumflex (LCx) arteries were measured in both cardiac phases for both approaches. Qualitative and quantitative analyses were performed in a blinded and randomized fashion. RESULTS: In volunteers, there was no significant difference in vessel sharpness between approaches (p > 0.05). In patients, analysis showed equal vessel sharpness for LAD and RCA (p > 0.05). LCx sharpness was greater with dNAV (p < 0.05). Visualized length with iNAV was 0.5 ± 0.4 cm greater than that with dNAV for LCx in diastole (p < 0.05), 1.0 ± 0.3 cm greater than dNAV for LAD in diastole (p < 0.05), and 0.8 ± 0.7 cm greater than dNAV for RCA in systole (p < 0.05). Qualitative scores were similar between modalities (p = 0.71). Mean iNAV scan time was 5:18 ± 2:12 min shorter than mean dNAV scan time in volunteers (p = 0.0001) and 3:16 ± 1:12 min shorter in patients (p = 0.0001). CONCLUSIONS: Image quality of iNAV and dNAV was similar with better distal vessel visualization with iNAV. iNAV acquisition time was significantly shorter. Complete cardiac diagnosis was achieved. Shortened acquisition time will improve clinical applicability and patient comfort.
No takes yet. Share an insight, caveat, or question.
Moyé et al. (2018) conducted a cross-sectional in Congenital heart disease (n=49). Image-based navigation (iNAV) for dual-phase 3D whole-heart MRI vs. Diaphragmatic navigation (dNAV) was evaluated on Mean scan time in patients (3:16 min shorter, p=0.0001). Image-based navigation for dual-phase 3D whole-heart MRI significantly reduced mean scan time by 3 minutes and 16 seconds compared to diaphragmatic navigation in patients with congenital heart disease, while maintaining similar image quality.
Synapse has enriched 10 related papers. Consider them for comparative context: