Answers are grounded in this paper and cite their sources.
A hydrostatic head viscometer and its novel viscosity equation were developed to determine flow characteristics of Newtonian and non-Newtonian fluids. The objective of this research is to test capabilities of the hydrostatic head viscometer and its novel non-Newtonian viscosity equation by characterizing rheological behaviors of well-known polyethylene oxide aqueous solutions as non-Newtonian fluids with 60 wt.% sucrose aqueous solution as a reference/calibration fluid. Non-Newtonian characteristics of 0.3–0.7 wt.% polyethylene oxide aqueous solutions were extensively investigated with the hydrostatic head viscometer and its non-Newtonian viscosity equation over a 294–306 K temperature range, a 0.14–40 Reynolds number range, and a 55–784 s−1 shear rate range at atmospheric pressure. Dynamic viscosity values of 60 wt.% sucrose aqueous solution were determined with the calibrated hydrostatic head viscometer and its Newtonian viscosity equation over a 3–5 Reynolds number range at 299.15 K and atmospheric pressure and compared with the literature dynamic viscosity value.
No takes yet. Share an insight, caveat, or question.
Kwon et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: