Key result
Axial-flow LVAD support induces reversible type 2 von Willebrand syndrome in ~100% of patients.
Cross-Sectional (n=26)
Answers are grounded in this paper and cite their sources.
BACKGROUND: Rotary blood pumps used as left ventricular assist devices (LVADs) allow for long-term support and may become suitable alternatives to heart transplantation. Effects of this technology on the coagulation system are not completely understood, leading to controversial anticoagulation protocols. Thus, we investigated the primary hemostasis in patients with chronic LVAD therapy. METHODS AND RESULTS: Twenty-six outpatients received axial flow LVAD (HeartMate II; Thoratec) for a median support time of 4.5 months. In a cross-sectional protocol, platelet aggregation in response to ADP and epinephrine, von Willebrand antigen (vWF:AG), and collagen-binding capacity (vWF:CB) were obtained. Von Willebrand factor (vWF) multimer analyses were performed, and patients were screened for bleeding events. This analysis was repeated after removal of the device for transplantation or recovery (n=12) and after a median of 15.5 months in ongoing patients (n=11). In all patients on devices, severe impairment of platelet aggregation as well as a loss of large vWF multimers were found. In 10 patients, a decreased vWF:CB/vWF:AG ratio was observed. Bleeding episodes occurred with an incidence of 0.17 per patient-year. After removal of the device, normal patterns of platelet aggregation, multimer analysis, and vWF:CB/vWF:AG ratio were recorded. In the second analysis of ongoing patients, impairment of platelet aggregation and loss of large vWF multimers were verified. CONCLUSIONS: A diagnosis of von Willebrand syndrome type 2 was established in all patients after LVAD implantation, and bleeding events confirmed this finding. Reversibility of this condition was found after removal of the device.
No takes yet. Share an insight, caveat, or question.
Meyer et al. (2010) conducted a cross-sectional in Acquired von Willebrand syndrome in patients with chronic left ventricular assist device therapy (n=26). Axial flow left ventricular assist device (HeartMate II; Thoratec) vs. Post-explantation state (after device removal for transplantation or recovery, n=12) was evaluated on Acquired von Willebrand syndrome type 2 (severe impairment of platelet aggregation and loss of large vWF multimers). Axial flow LVAD support caused reversible type 2 von Willebrand syndrome in 100% (26/26) of patients, with a bleeding incidence of 0.17 per patient-year.
Synapse has enriched 10 related papers. Consider them for comparative context: