Why the study?
Differences in the incidence and characteristics of silent cerebral events after radiofrequency AF ablation across different mapping catheters and ablation-guiding indices remained unclear.
Does the use of a mini-basket catheter with Rhythmia mapping increase the incidence of silent cerebral events compared to circular mapping with CARTO during atrial fibrillation ablation?
Does the use of a mini-basket catheter with Rhythmia mapping increase the incidence of silent cerebral events compared to circular mapping with CARTO during atrial fibrillation ablation?
The use of a mini-basket mapping catheter during atrial fibrillation ablation is associated with a significantly higher incidence of silent cerebral events compared to circular mapping catheters.
Mini-basket Rhythmia mapping was associated with more SCEs; leaves open whether this warrants technique changes pending randomized data.
INTRODUCTION: The difference in the incidence and characteristics of silent cerebral events (SCEs) after radiofrequency-based atrial fibrillation (AF) ablation between the different mapping catheters and indices used for guiding radiofrequency ablation remains unclear. This study aimed to compare the incidence and characteristics of postablation SCEs between the following two groups: Group C, Ablation Index-guided ablation using two circular mapping catheters with CARTO (Biosense Webster); Group R, local impedance-guided ablation using one mini-basket catheter and one circular mapping with Rhythmia (Boston Scientific). METHODS AND RESULTS: Of 211 consecutive patients who underwent an AF ablation and brain magnetic resonance (MR) imaging after the ablation, 120 patients (each group, n = 60) were selected by propensity score matching. SCEs were detected in 37 patients (30.8%). Group R had a higher incidence of SCEs (51.7% vs. 10.0%; p < .001) and more SCEs per patient (median, 3 vs. 1, p = .028) than Group C. A multivariate analysis demonstrated that nonparoxysmal AF and being Group R were independent positive predictors of SCEs (odds ratios, 6.930 and 15.464; both p < .001). On the follow-up MR imaging, all SCEs in Group C and 87.9% of the SCEs in Group R disappeared (p = .537). CONCLUSIONS: Group R had a significantly higher incidence of SCEs than Group C. Most probably the use of a complexly designed basket mapping catheter is the reason for the difference in the incidence of SCEs but further validation is needed. A nonparoxysmal form of AF may also increase the risk of SCEs during these ablation procedures.
No takes yet. Share an insight, caveat, or question.
Nakamura et al. (2020) studied this question.