Blood concentrations of essential trace elements can be used to diagnose conditions and diseases associated with excess or deficiency of these elements. Inductively coupled plasma mass spectrometry (ICP-MS), inductively coupled plasma atomic emission spectroscopy (ICP-AES), and graphite furnace atomic absorption spectrometry (GF-AAS) have been employed for such measurements, but maintenance and operation costs are high. X-ray fluorescence (XRF) detectability in cutaneous blood of iron (Fe), copper (Cu), zinc (Zn), and selenium (Se) was assessed as an alternative to ICP-MS. Three phantoms were made up of two polyoxymethylene (POM) plastic cylindrical cups of 0.6 mm and 1.0 mm thick walls and a 5.3 mm diameter POM cylindrical insert. Six aqueous solutions of Fe in 0 to 500 mg/L and Cu, Zn, and Se in 0 to 50 mg/L concentrations were poured into the phantoms to simulate X-ray attenuation of skin. Measurements using an integrated X-ray tube and polycapillary X-ray lens unit generated 24 calibration lines. Detection limit intervals in mg/L were (36–100), (14–40), (3.7–10), and (2.1–3.4) for Fe, Cu, Zn, and Se, respectively. Fe was the only element with detection limits lower than its 480 mg/L median human blood concentration. The estimated radiation dose and equivalent dose to skin were below those of common radiological procedures. Applications will require further instrumental development and finding a calibration method.