Why the study?
Does the random-zero sphygmomanometer yield systematically different blood pressure readings compared to the standard mercury sphygmomanometer?
Does the random-zero sphygmomanometer yield systematically different blood pressure readings compared to the standard mercury sphygmomanometer?
The random-zero sphygmomanometer yields systematically lower blood pressure readings than the standard mercury sphygmomanometer, indicating they should not be used interchangeably in epidemiologic studies.
Answers are grounded in this paper and cite their sources.
Random-zero sphygmomanometers may underestimate BP versus mercury at standard deflation; leaves open whether device-specific adjustments are required in legacy cohorts.
The random-zero and standard mercury sphygmomanometers are used frequently, and sometimes interchangeably, in epidemiologic studies. To determine whether there is a systematic difference between them, the authors measured systolic, fourth-phase, and fifth-phase diastolic blood pressures using both sphygmomanometers simultaneously in a series of six experiments. For most experiments, the system for simultaneous blood pressure measurements employed one cuff connected to both sphygmomanometers, which were carefully calibrated and read by two trained technicians using a double stethoscope. Order of use of the random-zero sphygmomanometer was randomly assigned, and technicians were blind to each others readings. At deflation rates of 2 mmHg/second, readings of the random-zero sphygmomanometer were systematically lower than those of the standard mercury sphygmomanometer: -0.9 mmHg for systolic, -1.8 mmHg for fourth-phase, and -1.8 mmHg for fifth-phase diastolic blood pressures (all differences, p less than 0.001). The difference persisted after adjustment for subject age, sex, heart rate, blood pressure level, observer, and room temperature, and was present for varying deflation rates. However, by draining the residual mercury from a random-zero sphygmomanometer and using the instrument as if it were a standard mercury sphygmomanometer, much of the difference between the two was eliminated. This suggests that the mechanism for the difference relates to the increased height of mercury in the random-zero manometer tube. The authors conclude that the random-zero and standard mercury sphygmomanometers should not be used interchangeably in epidemiologic studies.
No takes yet. Share an insight, caveat, or question.
Gaudemaris et al. (1985) studied this question.