Answers are grounded in this paper and cite their sources.
Solutions of vancomycin in phosphate-buffered saline, peritoneal dialysis effluent fluid and human serum were incubated at 37 degrees C for ten days and sampled at daily intervals. The samples were assayed for vancomycin content by a microbiological assay, HPLC and polarisation fluoroimmunoassay (Abbott TDX). The results obtained by HPLC and microbiological assay agreed well and indicated approximately 50% loss over ten days in serum and buffered saline and over 70% loss in dialysate. TDX results indicated losses of only 20% and 40%, respectively. Degradation products were prepared from vancomycin by acid hydrolysis and purified by HPLC. These purified products were shown to cross-react in the TDX assay. It is suggested that the TDX assay becomes non-specific in the presence of vancomycin breakdown products and thus overestimates true vancomycin concentrations.
No takes yet. Share an insight, caveat, or question.
White et al. (1988) studied this question.