Agreement was obtained within expected uncertainty limits between theoretically derived corrections and those derived experimentally. The correction factor for general recombination in continuous radiation is expected to be valid for dose rates up to at least 1000 cGy min -1 , which is unlikely to be exceeded in typical photon beams used for radiotherapy. For pulsed radiation it is expected to be valid for doses per pulse up to about 0.1 cGy, i.e. a correction of about 1.02. Although the corrections have been calculated for typical calibration factors for NE 2561 chambers, there is no need to amend the correction if a particular NE 2561 chamber has a somewhat different calibration factor. It occurs in the formulae in association with the assumed chamber volume, and the product of these two quantities is approximately constant for high-energy photon beams.