Why the study?
Renin plasma level determination is useful in diagnosing hypertension and monitoring therapy, differentiating primary hyperaldosteronism from renovascular hypertension.
Does the chemiluminescence immunoassay CLIA LIAISON® correlate with radioimmunoassay for measuring aldosterone and renin?
Does the chemiluminescence immunoassay CLIA LIAISON® correlate with radioimmunoassay for measuring aldosterone and renin?
The CLIA LIAISON method provides comparable results to traditional radioimmunoassay for measuring aldosterone and renin, supporting its clinical use in evaluating hypertension.
May support CLIA as non-radioactive alternative for aldosterone/renin testing; leaves open need for outcome validation before routine adoption.
Determination of renin plasma levels is useful in the diagnosis of hypertension and in the therapeutic follow-up of hypertensive patients. Plasmatic concentration of renin decreases in patients with hypertension due to a primary hyperaldosteronism, contrary to renovascular hypertension where concentrations of renin and aldosterone are both elevated. Blood samples (serum, EDTA plasma) were analysed using two different chemiluminiscent methods CLIA LIAISON® and radioimmunoassay for aldosterone (IMMUNOTECH Beckman Coulter) and renin (Cisbio Bioassay) measurements were compared. We used both methods to ascertain the correlation between serum vs. EDTA plasma levels of aldosterone (RIA, CLIA) and renin (IRMA, CLIA) and to compare aldosterone to renin ratios for CLIA and for radioimmunoassay: serum aldosterone to plasma renin and plasma aldosterone to plasma renin. We compared serum aldosterone CLIA vs. RIA (rP=0.933, P<0.001) and plasma renin determined using CLIA vs. IRMA (rP=0.965, P=0.062). Furthermore, we used both methods to establish the correlation between the serum vs. plasma levels of aldosterone: RIA (rP=0.980, P<0.001); CLIA (rP=0.994, P=0.353) and serum vs. plasma levels of renin: IRMA (rP=0.948, P<0.001); CLIA (rP=0.921, P=0.011). Aldosterone (serum, plasma) to plasmatic renin ratios for CLIA (rP=0.999, P=0.286) and for radioimmunoassay (rP=0.992, P=0.025). Our data demonstrate that renin and aldosterone concentrations obtained using CLIA correlate with renin and aldosterone concentrations using radioimmunoassay methods. Correlation coefficients of pair results ranged from 0.921 to 0.994. Aldosterone (serum, EDTA plasma) to plasmatic renin ratios are comparable and any of them can be used with no significant differences found.
No takes yet. Share an insight, caveat, or question.
Uhrová et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: