Key result
Cardiac catheterization confers ~28 R patient skin exposure alongside operator eye and gonadal radiation.
Observational
Answers are grounded in this paper and cite their sources.
A survey of the literature was conducted to establish the anticipated ranges of exposure to both physicians and patients during cardiac catheterization. A brief explanation of a technic for using time-lapse photography and a computer model for exposure calculation is presented. The thermoluminescent dosimeter (TLD) results used as controls for exposure values calculated by the developed technique are presented in detail. Physician eye exposures of approximately 20 mR per cardiac catheterization procedure were measured, which would suggest a limit of five procedures per week for physicians. The average patient skin entrance exposure of 28 R is high, as is the 12 mR gonadal exposure; however, they are accepted because of the possible benefits of the procedure.
No takes yet. Share an insight, caveat, or question.
F G Rueter (1978) conducted an observational in Cardiac catheterization. Radiation exposure during cardiac catheterization vs. Thermoluminescent dosimeter (TLD) controls was evaluated on Physician eye exposure, patient skin entrance exposure, and patient gonadal exposure per procedure. Cardiac catheterization yielded physician eye exposures of ~20 mR, patient skin entrance exposures of 28 R, and patient gonadal exposures of 12 mR per procedure.
Synapse has enriched 10 related papers. Consider them for comparative context: