Why the study?
Does device-based atrial pacing terminate spontaneous atrial tachyarrhythmias in patients with ventricular arrhythmias receiving a dual-chamber ICD?
Does device-based atrial pacing terminate spontaneous atrial tachyarrhythmias in patients with ventricular arrhythmias receiving a dual-chamber ICD?
Device-based atrial pacing can successfully terminate nearly half of spontaneous atrial tachyarrhythmia episodes in patients with ventricular arrhythmias and a dual-chamber ICD.
Answers are grounded in this paper and cite their sources.
Supports atrial pacing for AT/AF termination in dual-chamber ICD patients; hypothesis-generating without randomized outcome data.
BACKGROUND: Although overdrive pacing for treating atrial flutter is well established, the efficacy of device-based atrial pacing for treating spontaneous atrial tachyarrhythmias in patients with implantable cardioverter defibrillators (ICD) is unknown. This study evaluated the efficacy of novel pacing therapies for treating atrial tachyarrhythmias in patients receiving a dual-chamber ICD to treat ventricular tachyarrhythmias. METHODS AND RESULTS: A Jewel AF ICD was implanted in 537 patients with ventricular arrhythmia who were followed for 11.4+/-8.2 months (74% had a documented history of atrial tachyarrhythmias). The device discriminated atrial tachycardia (AT) from atrial fibrillation (AF) on the basis of cycle length and regularity, and it used 3 different methods of overdrive atrial pacing (Ramp, Burst+, and 50-Hz burst) to treat AT episodes and one method (50-Hz burst) to treat AF episodes. Pacing successfully terminated 59% of 1500 spontaneous AT episodes in 127 patients and 30% of 880 AF episodes in 101 patients (P<0.001). With AT and AF episodes combined, pacing efficacy was 48%. Pacing efficacy was significantly reduced at AT cycle lengths </=220 ms and AF cycle lengths </=160 ms (P<0.01) The median time from pacing to AT/AF termination was 1.1 minute for the pacing successes and 96 minutes for the failures (P<0.001). CONCLUSIONS: Atrial pacing terminated 48% of AT/AF episodes in patients with a history of ventricular arrhythmias. Pacing efficacy was greater for device-classified AT than AF. Pacing efficacy was significantly influenced by tachycardia cycle length. Successful pacing significantly reduces the time required for AT/AF termination.
No takes yet. Share an insight, caveat, or question.
Adler et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: