A planar retarding‐potential analyzer will be included in each of the Atmosphere Explorer (AE) payloads. The primary functions of the instrument are to supply ion‐temperature and ion‐concentration data, which it will do at least once every 40 km of flight path in the region of interest with an expected accuracy of better than 2%. In addition, the instrument will determine ion‐drift velocities and energy spectra of both the thermal and suprathermal electrons. During eclipse, negative‐ion concentrations greater than 0.3 cm −3 can be detected, though the normal positive‐ion sensitivity is approximately 1.5 ions cm −3 . Changes in the ion concentration along the flight path greater than 0.1% can be monitored with a spatial resolution of less than 40 m.
No takes yet. Share an insight, caveat, or question.
Hanson et al. (1973) studied this question.