We demonstrate a high data-rate light-pulse atom interferometer for measuring acceleration. The device is optimized to operate at rates between 50 Hz to 330 Hz with sensitivities of 0.57μg/Hz to 36.7μg/Hz, respectively. Our method offers a dramatic increase in data rate and demonstrates a path to applications in highly dynamic environments. The performance of the device can largely be attributed to the high recapture efficiency of atoms from one interferometer measurement cycle to another.
No takes yet. Share an insight, caveat, or question.
A 2012 study studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: